To prepare your samplescell sample, please follow the instructions below:

  • Equilibrate the LUTHOR scd Buffer buffer at RT for 10min before diluting the cell sample

  • Filter the cell sample with the cell strainer (at least 1ml) to prevent clogging of the LUTHOR scd dispensing tipTip

  • Check cell concentration and viability using a cell counter (with Trypan Blue staining, Dual Acridine Orange (AO) staining or Propidium Iodide (PI) staining), fluorescent cell counter or microscope. Evaluate the % of cell debris and cell aggregates to set-up the thresholds for cell counting during dispense.

NOTE: Cell viability should be at least 90% to ensure best results for with the LUTHOR HD library preparation.

  • Dilute the sample ideally to 150K cells/ml

  • Add the LUTHOR scd buffer to the sample and pipet slowly for mixing (by using the wide-bore filter tips provided with the LUTHOR Single Cell single cell dispenser)

  • Place the sample on the hold tube rack.

IMPORTANT: Keep your cells on ice until dispensing!