The Particle Threshold corresponds to the minimum impedance signal defined for a cell which helps to distinguish particles of interest from debris and other non-relevant particles.Below this value, particles are not detected during dispensing. A default threshold of 200 ohms is set-up in the analysis software (threshold defined for mammalian cells >10µm). This value can of course be adjusted depending on the experimental conditions (cell line and medium used).

...

Additionally, thresholds can also be adjusted based on the sample distribution (see FAQ: How do I evaluate the concentration of my sample ?).

The distribution of the sample should follow a gaussian curve. Events detected before or after the curve can be considered as debris or maybe non-relevant particles.

...