Anchor_Hlk77025319_Hlk77025319 Anchor_Hlk65531786_Hlk65531786The kit uses total RNA as input, hence no The recommended input range is 1 - 500 ng of total RNA.

No prior poly(A) enrichment or rRNA depletion of total RNA is required. Low quality RNA including formalin-fixed parrafin embedded (FFPE) samples can also be used for QuantSeq.

The ideal amount of total RNA needed for QuantSeq depends on the poly(A) RNA content of the sample in question. This The protocol was tested extensively with various cell cultures, animal and plant tissues, yeast, fungi, drosophila, and human reference RNA.


As a starting point, we recommend performing the protocol initially with 500 ng total RNA. RNA inputs ≥200 ng are recommended to detect low abundant transcripts efficiently.
The minimum recommended input amount is 1 ng for high quality total RNA (down to 100 pg is possible). Protocol modifications apply for low input (≤ 10 ng) and low quality or FFPE RNA (see FAQ 4 below, and Appendix B and C of the QuantSeq User Guides).
Anchor_1.4_What_is_1.4_What_isWhat is the minimum input amount? Are there any recommendations for low RNA input?
Anchor_Hlk68031751_Hlk68031751 Anchor_Hlk77025674_Hlk77025674When using <10 ng or <1 ng of total RNA input, please follow the recommendations indicated in the table below (also noted in Appendix B of the QuantSeq User Guides):
Image Removed