Do my QuantSeq FWD-UMI reads look okay?

Lexogen Online FAQs

Do my QuantSeq FWD-UMI reads look okay?

QuantSeq FWD-UMI libraries contain a 4 nt spacer that separates the 6 nt UMI at the start of the read from the random priming sequence that initiates second strand synthesis. The spacer has the same sequence (TATA) in all library fragments, which generates low diversity signals for cycles 7-10 of Read 1. This low diversity signature is visible in both sequencing run results (e.g., SAV files, or basespace run results) and FastQC analysis results.

The figures below show the FastQC per base sequence content for QuantSeq FWD (standard, no UMI, Fig. 1) and QuantSeq FWD-UMI (Figure 2) libraries.
We recommend adding a minimum of 5 – 15 % PhiX (maximum 30 %) spike-in when sequencing QuantSeq FWD-UMI libraries on all Illumina instruments. The higher PhiX percentage corrects for the low diversity of the spacer.


ATTENTION! The optimal PhiX spike-in percentage may differ between instruments. See also: Do you have recommended loading concentrations for Lexogen libraries on Illumina sequencing instruments?


Figure 1 | FastQC Per Base Sequence Content profile for QuantSeq FWD library pool sequenced on NextSeq 500 (no UMIs).

Figure 2 | FastQC Per Base Sequence Content profile for QuantSeq FWD-UMI library pool sequenced on NextSeq 500.


NOTE: The QuantSeq FWD-UMI data analysis pipeline provided by Lexogen trims the UMI and spacer from the beginning of Read 1 before read Quality Control is performed. Therefore, FastQC reports for raw and trimmed reads look the same as for Standard QuantSeq FWD libraries (without UMIs, see https://lexogen-customerportal.atlassian.net/l/c/eWYn0JSt for pipeline details).


Related content

How are UMIs read-out during sequencing? What read length is needed?
How are UMIs read-out during sequencing? What read length is needed?
More like this
How do I analyze my QuantSeq FWD-UMI Sequencing Data?
How do I analyze my QuantSeq FWD-UMI Sequencing Data?
More like this
What sequencing read length is best suited for QuantSeq FWD libraries ?
What sequencing read length is best suited for QuantSeq FWD libraries ?
More like this
What loading amount is recommended for QuantSeq FWD-UMI libraries?
What loading amount is recommended for QuantSeq FWD-UMI libraries?
More like this
Are Unique Molecular Identifiers (UMIs) included in QuantSeq FFPE?
Are Unique Molecular Identifiers (UMIs) included in QuantSeq FFPE?
More like this
Why do I see a high % of Duplicate Sequences (FastQC) in my QuantSeq data?
Why do I see a high % of Duplicate Sequences (FastQC) in my QuantSeq data?
More like this

Have feedback for us or need more information? Send us a request or write to us at support@lexogen.com.