The impedance signal of a single cell should appear as a single and isolated peak.

 

image-20240816-134141.png

 

At the start, variations in the impedance signal are observed, primarily due to differences in conductivity between the buffer inside the LUTHOR scd Tip and the buffer in the well (points 1, 2 and 3). When dispensing begins, the baseline remains continuous until a particle passes through the tip and is detected . This event is marked by a distinct and well-defined peak (point 4). After dispensing, fluctuations in the impedance signal can be seen as the tip exits the well (points 5 and 6).

When a doublet is detected, the impedance signal displays two peaks which are close in proximity to one another. An example of a doublet can be found below:

image-20240826-024643.png

Particles/debris smaller than a cell can also be detected and would result in small ohm values (see red arrows in Figure below):

image-20240820-130637.png

 