Lexogen Online FAQs

Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 3 Current »


Many liquid handlers do not have additional heat block positions on deck to pre-warm mastermixes. However, the most critical step for QuantSeq first strand cDNA synthesis is that the RNA / FS1 samples are kept continuously at 42 °C after denaturing (step 2), and while the FS2 / E1 mastermix is added. Although pre-warming this mastermix helps to prevent mishybridization by maintaining the sample temperature of the reactions after denaturing, the pre-warming can be omitted for automated protocols as long as the FS2 / E1 mastermix is prepared in advance and kept at room temperature on deck. This mastermix should not be cooled or kept on a refrigerated block on deck prior to adding to the samples (at step 3).

  • No labels